UF-100尿沉渣分析仪与相差显微镜在血尿来源鉴别中的应用 被引量:14

Supplementary UF-100 and Phase-contrast Microscopy Examination to Define the Origin of Hematuria
下载PDF
导出
摘要 目的评价UF-100尿沉渣分析仪与相差显微镜在血尿来源鉴别中的价值。方法选取肾病内科住院病人血尿患者102人,分别用UF-100尿沉渣分析仪与相差显微镜检查分析尿液中红细胞的相关参数。结果对于血尿来源,UF-100尿沉渣分析仪检查的敏感性(93.1%)高于相差显微镜检查的敏感性(83.3%),P<0.05;而相差显微镜检查的特异性(80.0%)高于UF-100尿沉渣分析仪检查的特异性(53.3%),P<0.05。结论UF-100尿沉渣分析仪与相差显微镜结合使用是血尿来源鉴别的最好策略。 Objective To evaluate the clinical usefulness of UF-100 and phase-contrast microscopy examination of urinary sedimentation in identification of the origin of hematuria. Methods Urinary samples from 102 inpatients with kidney disease were examined by UF-100 and phase-contrast microscopy respectively. Results The sensitivity of identification of origin UF-100 was 93.1%, which was significantly higher than that 83.3% of the phase-contrast microscopy (P< 0.05 ); while the speciality of UF-100 was only 53.3% that was significantly lower than that 80.0% of the phase-contrast microscopy (P<0.05). Conclusion Supplementary UF-100 and phase-contrast microscopy for the localization of the origin of hematuria raise the sensitivity and specificity.
出处 《实用预防医学》 CAS 2005年第4期791-793,共3页 Practical Preventive Medicine
关键词 血尿 UF-100 相差显微镜 尿液 红细胞 Hematuria Sysmex UF-100 automated urinalysis analyzer Phase-contrast microscopy Urine Erythrocyte
  • 相关文献

参考文献9

  • 1Thomas J, Pillsworth JR, Virginia M, et al. Differentiation of Renal from Non - Renal Hematuria by Microscropic Examination of Erythrocytes in urine[J]. Clin Chem, 1987,33(10): 1791 - 1795.
  • 2马骏龙,丛玉隆,陈淑云,周建山,丁颖.全自动尿沉渣分析仪检测尿红细胞鉴别血尿来源的探讨[J].中华医学检验杂志,1997,20(5):305-305. 被引量:54
  • 3金大鸣.UF-100型尿沉渣全自动检测仪评价[J].中华检验医学杂志,1998,22(3):61-62. 被引量:145
  • 4Hyodo E, King H. Automated Methods in Urinalysis[J]. Clin Lab Sci,1998, 11(1):44-46.
  • 5李幼姬 叶任高 等.诊断肾小球血尿的一种简便方法[J].中华内科杂志,1984,23(2):99-99.
  • 6熊立凡,顾晓菁,董晓蓓.尿棘形红细胞在肾性和非肾性血尿鉴别中的意义[J].上海医学检验杂志,2000,15(2):74-76. 被引量:25
  • 7ECLM. European Urinalysis Guidelines[M]. In: Kouri T, Fogazzi G,Gant V, Hallander H, Hofmann W, Guder WG, editors, Scand J Clin Lab Invest 2000(suppl).
  • 8Product Development Division, TOA Medical Electronics. Performance of the Sysmex UF- 100 fully automated urine cell analyzer[J]. Sysmex JInt, 1996,6:36-40.
  • 9Fenili D. UF- 100 Evaluation Experience[M]. In: The Sysmex Urine Flow Cytometry Workshop Proceedings, Hamburg: Sysmex Europe GmbH, 1997, PP. 33 - 40.