摘要
为进一步解决胸腹部MR成像中呼吸运动伪影的干扰,通过呼吸频率与重复时间(TR)相互关系的分析,认为由呼吸频率修正TR值与呼吸门控所起的作用类似,且当呼吸门控与修正TR值同时使用时,更有助于呼吸门控功能的发挥,并推导出关系式:TR=×1000。经30例成像效果观察并予初步探讨。
In order to reduce the artifacts caused by respiratory movement in thoracoab-dominal MRI, the relation between respiration frequency and repetition time was analyzed. The re-sults showed that the role of respiration frequency corrected TR value was similar to that of respira-tory gating. Moreover,with combined use of respiratory gating and corrected TR value,the function of respiratory gating will be further facilitated.The following formula was used:TR=×1000.30 cases were used for demonstration.
出处
《中华放射学杂志》
CAS
CSCD
北大核心
1995年第3期165-167,共3页
Chinese Journal of Radiology