摘要
建立了测定化探样品中微量溴的新方法,即热水解-流动注射分析方法,并成功地用于化探样品中溴的测定。溴的热解率为98%以上;精密度为3%~5%,热解后样品溶液可直接用于流动注射分析测定,进样频率为180次/h精密度(RSD%)为0.24%(n=8,对含溴离子为0.5ug/mIBr  ̄-样品溶液而言);检出限为0.05ug/mlBr ̄-;线性范围为0.05~1.0ug/mlBr ̄-。用本法对3个地球化学标准参考样(GSS-1,GSS-3,GSS-7)进行分析测定,其测定结果与这些标准样品的推荐值非常地一致。
A new method for determination of bromide in geochemical exploration samples bymeans of heat water relese with flow injection analysis is established in this paper.Meanuhile, the method has been successfully applied to determine the bromide in geochemi-cal exploration samples. The recovery ratio of bromide is more than 98%,The precision isabout 3% 5%. The sample solution can be directly determined by fiow injection analysis.The frequency for determining samples comes to 180 times per hour. The precision(RSD,%)is 0.24%(n=8,for samples containing 0.5ug/mI Br ̄-).The detection limit of Br ̄-is 0.05ug/ml.Linear calibration ranges between 0.05 and1.0ug/mI Br ̄-. 3 geochemical ston-dard reference samples(GSS 1 ,GSS 3, GSS 7)have been dettrmined by this method.Their data are quite identical to the recommended values of the samples.
出处
《矿产与地质》
1995年第1期69-72,F003,共5页
Mineral Resources and Geology
关键词
溴
热解
流动注射分析
地球化学勘探
矿样