期刊文献+

Humphrey视野分析仪检查结果的分析和判断

Analysis and Judgement of the Results of Humphrey field Analyzer
下载PDF
导出
摘要 自动视野计正在广泛地用于眼科临床,使检查结果更趋向客观,可靠,为定量分析视功能提供了较精确的数据。那么,采用什么方法、根据什么标准分析和判断自动视野计的检查结果?围绕这个问题。 The study was intended to find the method and criteria for analysing and judging the results of the automatic perimeter according to the defect. Based on the control study, we think that the criteria for the early glaucoma visual field should be: (1) more than two adjacent points over 5 dB, or single point over 6 dB within 10 degree; (2) more than three adjacent points over 6 dB, or single point over 8 dB between 10-20 degree. The careful consideration must be given to the change of threshold within the region beyond 20 degree. The sensitivity and specificity for the criteria is 100% and 90%, respectively.
出处 《眼科研究》 CSCD 1989年第3期152-154,共3页 Chinese Ophthalmic Research
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部