摘要
本文研究了Fe^(3-)的单点络合滴定法,导出了定量关系式,比较了以铂电极为指示电极和以 PVC膜四氯络铁(Ⅲ)阴离子电极为指示电极时电位测量的稳定性及所得的分析结果.在Fe^(3-)浓度为1.00×10^(-3)mol/L时.18次实验的平均值为1.00×10^(-3)mol/L,相对偏差为0.2%;Fe^(3-)的浓度为1.00×10^(-4)mol/L时,15次测定的平均值为及1.02×10^(-4)mol/L,相对偏差小于1.7%;Fe^(3+)在5.0×10^(-5)mol/以上时,相对偏差小于0.1%;其浓度小于10^(-3)mol/L而大于5×10^(-4)mol/L时,相对偏差小于0.5%,且精密度好,该法分析 Fe^(3+)简单,快速,精密度高.准确度好,可实现无汞,无K_2Cr_2O_7,定铁,并可自动化.
The method for the complexometric titration of iron using EDTA as complexing a-gent was studied, and the quantitative equation is given. Pt electrode and PVC-based membrane tetrachloroferrate ( Ⅲ ) ion-selective electrode were used as indicating electrode. The stabilization of the electrode potential of the two electrodes as well as the results of the analysis were compared. At [Fe3+]=1. 0×10×3 mol/L, the average value of 18 determination is 1. 00×10-3 mol/L, relative error is 0. 2%. At Fe3+ concentration 1. 00×10-4 mol/L, the average value of 15 determinations is 1. 02× 10-4 mol/L, relative error is 1. 7%. The analytical results are very satisfactory. The simple and fast method can be automated, and avoid the use of mercuric chloride and potassium dichromate.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
1995年第9期1043-1046,共4页
Chinese Journal of Analytical Chemistry
基金
湖南省教委优秀青年骨干教师基金