摘要
采用阳离子交换树脂离子分离铜,发生氢化物原子吸收光谱法测定硒。上柱溶液酸度在0~7%,硒的回收率在96~101%,过柱溶液中铜的浓度低于0.01μg/mL。方法检出限为5×10 ̄(-5)%,分析紫铜中4.4×10 ̄(-4)%的硒,相对标准偏差为3%。
method was established for determi-nation of Se in copper by ion exchange sepa-ration hydride generation atomic absorptionspectuometry. The recovery is 96~101%andthe Concentution of residual Cu is Low as0.01%μg/mL in eluate ,The method has beenapplied to the determination of Se Concen-tration of 4.4 × 10 ̄(-4)%,the RSD is 3%withdete tion Limit is 5 × 10 ̄(-5)%
出处
《分析试验室》
CAS
CSCD
北大核心
1995年第6期49-52,共4页
Chinese Journal of Analysis Laboratory
关键词
原子吸收法
硒
紫铜
Hydride generation AAS
ion ex-change separation