摘要
在吸收型钠离子光化学传感器的敏感膜中加入合适的荧光试剂,应用荧光内滤效应研制成的荧光增强型光纤传感器,在测量灵敏度和抗背景干扰能力方面均有较大的改善,对血清和矿泉水样品中的钠离子含量进行了分析,获得了满意的结果。
Recently, many reversible optical chemical sensors based on ion-exchange schemes for sensing metal ions with comparable selectivities are developed. Most sensors of this kind reported so far are based on absorbance measurements. However, luminescence technique is an especially suitable optical chemical sensing method for designing these sen sors. This is due in part to its inherent sensitivity, its almost complete inertness to turbidity effects, and its flexibility with respect to geometric arrangements. In this paper, a fibre-optic sensor of fluorescence enhancement for sodium ions has been developed on the basis of ion-exchange sensing scheme. A suitable fluorescer is incorporated in the sensing membrane containing a lipophilic pH-indicator and sodium-selective neutral ionophore. Fluorescence varies as a function of the varying absorption of the pH-indicator which in turn is modulated by the sodium concentration through the inner filter effect of fluorescence. The sensor shows a linear reversible response to sodium ion over the range of 1× 10 ̄(-2)─2×10 ̄(-5) mol/L with a detection limit of 8 × 10 ̄(-6)mo/L. The response time of the sensor is about 2 ̄3min. The concentrations of sodium ions in serum and mineral water samples are determined by this sensor with satisfactory results.
出处
《高等学校化学学报》
SCIE
EI
CAS
CSCD
北大核心
1995年第4期518-522,共5页
Chemical Journal of Chinese Universities
基金
国家教育委员会优秀年轻教师基金
关键词
荧光内滤效应
钠离子
光化学传感器
Inner filter effect of fluorescence, Sodium ion, Optical chemical sensor, Optode.