期刊文献+

用AFS-930型双道原子荧光光度计同时测定水样中的砷和汞 被引量:6

Simultaneous Mensuration of Arsenic and Mercury in Water Using AFS-930 Atomic Fluorescence Spectrometer
下载PDF
导出
摘要 应用AFS-930型双道原子荧光光度计同时测定水中As和Hg,操作方法简单快速,基体干扰少.砷和汞的检出限分别为As:0.05μg/L,Hg:0.02μg/L;相对标准偏差(RSD)为As≤3.11%,Hg≤6.94%;回收率As在96.1%~106.5%之间,Hg在97.5%~107.0%之间;相关系数As为0.9999,Hg为0.9994. The operation of the AFS-930 atomic fluorescence spectrometer for simultaneously mensurating As and Hg in the water was simple and rapid, with little interference by base. The detection limits for arsenic and mercury were 0. 05μg/L and 0. 02μg/L respectively. The relative standard deviation for As was less than 3.11% , while 6.94% for Hg. The recovery for As was in the rang of 96.1% - 106.5% , and 97.5% - 107.0% for Hg. And the As correlation coefficient was 0. 9999, while the Hg was 0. 9994.
出处 《广东化工》 CAS 2005年第7期61-62,共2页 Guangdong Chemical Industry
关键词 氢化物发生 原子荧光 同时测定 双道原子荧光光度计 930型 水样 相对标准偏差 hydride generation atomic fluorescence water arsenic mercury simultaneous determination
  • 相关文献

参考文献4

二级参考文献8

  • 1北京海光仪器公司.原子荧光光度分析手册[M].北京:化学工业出版社,1995..
  • 2郭小伟 杨密云.氢化物发生无色散原子荧光光谱法在分析中的应用[J].分析化学,1980,8(5):466-466.
  • 3强卫国 何照明.氢化物发生原子荧光光度法测定食品中砷[J].环境与健康,1992,:32-32.
  • 4[2]魏复盛等,编著.水和废水监测分析方法指南[M].(下册).北京:中国环境科学出版社,1998,12~25.
  • 5(GB/T8538 1995)饮用天然矿泉水检验方法,1996年,77页
  • 6北京海光仪器公司,原子荧光光度分析手册,1995年
  • 7强卫国,环境与健康杂志,1992年,增刊,32页
  • 8郭小伟,分析化学,1980年,8卷,5期,466页