硫代巴比妥酸测定脂质过氧化物方法的改进 被引量:6

IMPKOVEMENT OF MEASUREMENT OF LIPID PEROXIDE BY THIOBARBITURIC
下载PDF
导出
摘要 为提高硫代巴比妥酸(TBA)比色法测定红细胞过氧化物的灵敏度,对该法的实验过程进行了两项改进:用50mmol/L的亚硫酸钠代替水作为溶血剂;用20%三氯乙酸代替冰醋酸作反应的酸性介质。实验结果表明吸光度可提高2.8倍,在2.5~50umol/L内线性良好,日内重复性实验x±s、CV分别为21.78±0.69nmol/gHb和3.17%,日间重复性实验x±s、CV分别为20.17±0.52nmol/gHb和2.58%,低、高两管浓度的平均回收率为98.25%和101%。60例正常参考值27.17±4. The laboratory method has been reformed for the purpose of improvingthe sensitivity of measurement of erythrocyte peroxide with thiobarbituric acid by colorimet-ric method. 50 mmol/L sodium suifite,instead of water,was used for hemolytic agent and20% trichloroacetic acid , instead of glancial acetic acid, was the acid medium, The resultshowed a good filiform within 25~50 umol/L and 2.8 times increase in absorbance, The val- ue of x±s and CV obtained by repeated test within one day were 21. 78±0. 69 nmol/g Hband 3.17%respectively, and within daytime were 20. 17±0. 25 nmoL/g Hb and 2. 58%,Theaverage rate of recovery for low and high tubes density were repectively 98.25%and 101%. The normal reference value of the sixty cases was 27. 71±4.32 nmol/g Hb.
作者 杨伟宗 刘明
出处 《九江医学》 1995年第1期25-27,共3页 Jiujiang Medical Journal
关键词 硫代巴比妥酸 脂质过氧化物 临床检验 thiobarbitutic acid sodium sulfite trichloroacetic acid lipid peroxide erythrocyte