期刊文献+

流动注射分光光度法测定铁的研究 被引量:5

The Spectrophotometric Determination of Iron(Ⅲ)in Flow Injection Analysis
下载PDF
导出
摘要 本文利用反向流动注射分析法(rFIA)对Fe(Ⅱ)-CAS-TritonX-100-HDMAA显色体系进行了显色条件、流路参数等的优化研究。测得采样频率为90次/小时,体系的测量波长为660nm,测量的线性范围为0~1.6μg/mlFe(Ⅱ),检测限为0.002μg/ml。用本法测定了水质标样、油田注水、自来水样中的痕量铁,结果令人满意。 Colour conditions and flow path parameters have been optimized in Fe(Ⅲ)-CAS-Triton X-100-HDMAA system by reverse flow injection analysis(rFIA). The measuringwavelength is 660 nm. The linear range is 0-0.6 μg/ml iron(Ⅲ)with a detection limit of0.002μg/ml injection rate is 90/h. The amount of the total iron in standard water sample,the tap water in Beijng and the water filled in the oil well in Daqing oil field has been deter-mined with satisfactory results.
作者 周磊
出处 《山东轻工业学院学报(自然科学版)》 CAS 1995年第3期40-45,共6页 Journal of Shandong Polytechnic University
关键词 反向流动注射 分光光度法 测定 everse flow injection analysis,spectrophotometry, iron