手持式非接触眼压计测量值准确性的比较 被引量:1

Evaluation of Accuracy of Measuring Intraocular Pressure by Handheld Non-contact Applanation Tonometer
下载PDF
导出
摘要 目的:对手持式非接触眼压计测量眼压的准确性及可靠性进行评价。方法:分别采用keeler,手持非接触眼压计和R900,Goldmann压平眼压计对58例患者(113只眼)的限内压进行了测量并对两种测量方法的测量结果进行了比较。结果:采用手持非接触眼压计测得的平均眼压为16.31±5.59mmHg(1mmHg=0.1333kpa),采用Goldmann压平眼压计测得的平均眼压值为17.49±6.13mmHg。经统计学t检验,两种测量结果差别无统计学意义(P>0.05)。经直线相关和回归分析发现两种方法测量值间存在正相关(r=0.8942,b=0.8154)。结论:手持非接触眼压计同Goldmann压平眼压计相比具有同样的精确性和可靠性,可用于青光眼临床和青光眼筛选。眼科学报 1995;11:86—88。 Purpose: To evaluate the accuracy of measuring intraocular pressure by handheld non-contact applanation tonometer.Method:58 patients' (113 eyes) intraocular pressure were measured by Keeler, non-contact tonometer and R 900 Goldmann applanation tonometer and the results of measurement of intraocular pressure by the two kinds of tonometers were compared.Result: The mean intraocular pressure measured by non-contact is 1 6 . 31 ± 5. 59mmHg and 17. 49±6. 13mmHg(1mmHg = 0. 1333kPa) by Goldmann applanation tonometer, respectively. There was no statistical significance to be found (P>0. 05) between the two methods. By linear correlation and regression analysis, a positive correlation was found between the two methods (r = 0. 8942, b = 0.8154).Conclusion : The handheld non-contact tonometer has the same accuracy and reliability of measurement of intraocular pressure comparing with Goldmann applanation tonometer, and it can be used in glaucoma clinic and screening. Eye Science 1995; 11:86-88.
出处 《眼科学报》 1995年第2期86-88,共3页 Eye Science
关键词 眼压测量 眼压计 测量值 准确性 glaucoma intraocular pressure
  • 相关文献

同被引文献3

  • 1赵家良 等.-[J].中华眼科杂志,1986,22:10-10.
  • 2陈秀琦,眼科学报,1995年,11卷,86页
  • 3赵家良,中华眼科杂志,1986年,22卷,10页

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部