摘要
介绍了在4π高气压正比计数器与碘化钠闪烁计数器符合装置上测量电子俘获核素 ̄(57)Co的效率外推法。测定了不同充气压力和不同工作电压对测量结果的影响。衰变率的合成不确定度为0.4%。这一方法也可用于具有类似衰变图的低Z电子俘获核素的测量。
An efficiency extrapolation method was used to measure the activity of ̄(57)Co by the 4πβ(PPC)-γ coincidence system,and the effects of pressure and voltage on the measurement were de-scribed in detail.The method also can be used to measure the activity of the some of nuclids withsimilar decay;The total uncertainty is estimated to be about 0.4%.
出处
《原子能科学技术》
EI
CAS
CSCD
北大核心
1995年第2期120-125,共6页
Atomic Energy Science and Technology
关键词
正比计数器
电子俘获
核素
外推法
钴57
pressure proportional counter Electron capture nuclide Efficiency extra-polation