摘要
目的 BD模拟测试对灭菌器的监测。方法通过使用不同的BD模拟测试方法 ,监测测漏保压测试失败的灭菌器,观察各测试装置的结果进行比较。结果通过监测与运行比较发现,在不同状态漏气的情况下,各BD模拟装置的监测结果各不一样。结论 PCD装置的灵敏性高于一次性BD包。
Objective Monitoring with BD simulation tests for sterilizer.Methods Monitoring failure sterilizer in holding pressure leak test with different BD simulation tests,then observe the results of the test device and make comparison.Results In the case of different state of leakage,all monitoring results of BD analog devices compared by the running of monitoring are different.Conclusion The sensitivity of PCD device is higher than one package of BD by comparison.