电极的校正及应用

CALIBRATION AND APPLICATION OF COATED WIRE CALCIUM ION SELECTIVE ELECTRODE
下载PDF
导出
摘要 本文应用EDTA作为钙离子缓冲溶液,对自制涂丝钙离子选择电极(Ca-CWISE)进行了校正,与用标准系列稀释法相比,电极的线性范围扩大了三个数量级。文章还利用自制Ca-CWISE,采用Thomas等所用的简易pH-pM法,测定了柠檬酸钙的浓度稳定常数,IgK_(CaCit^-)=3.86(30℃,I=0.1)。 This paper describes that calcium ion beffers, EDTA, is used for calibration of coated wire calcium ion selective electrodes made in our laboratory. Compared with the calibration by serial dilution standards, the working ranges of coated wire calcium ion selective electrode have been extended over three orders of mag- nitude. Detection limit is about 2×10^(-8) mol/L The concentration formation constant of calcium citrate complex was deter- mined by this electrode, i. e. lgK_(CaCi)=3. 86(30℃, I=0. 1) .
出处 《中南矿冶学院学报》 CSCD 1989年第5期576-580,共5页
关键词 电极 校正 电化学分析 稳定常数 ion elective electrode calibration of electrode determination of stability constants