食管曲张静脉测压数字图像系统的研究 被引量:3

Study of a digital video system for manometry of esophageal varices
原文传递
导出
摘要 目的观察食管曲张静脉测压数字图像系统的可靠性。方法制备食管及食管内曲张静脉模型,采用食管曲张静脉测压数字图像系统对5例肝硬化食管静脉曲张患者进行测压。食管曲张静脉测压数字图像系统由测压仪、压力传感器、视频采集卡及专用计算机程序组成。结果不同直径(3、6、8mm)模拟食管曲张静脉测压值与实际压力值间均呈良好相关性(r>0.975,P<0.01),未发现明显的测压值偏差(偏差的95%可信区间上下限介于-2.1~0.3cmH2O)。5例患者的食管曲张静脉压力的气囊测压值与该系统测得的压力值有较好的相关性(r=0.999,P<0.01)。结论食管曲张静脉测压数字图像系统能准确地测量食管曲张静脉压力。 Objective To develop a digital video system for esophageal variceal manometry and to evaluate its reliability. Methods The system consists of esophageal variceal manometer, pressure transducer, video capture card and special computer programs. An artificial esophagus containing water-filled latex tube was designed, in which the pressure had been measured by this system in vitro. The pressures of esophageal varices had been measured by using the method in five patients. Results In vitro study, a good correlation was showed between the actual pressure and measured pressure in the artificial varices with different diameter (3,6,9 mm)(r=0. 975, P〈0.01). No obvious measurement bias was found for any of the varices (95% confidence interval of bias: -2.1 cm H2O to 0.3 cm H2O) . The measurement in five patients also showed good correlation between the pressure measured by this method and the pressure measured by the balloon technique(r=0. 999, P〈0. 01). Conclusions The esophageal variceal pressure can be measured accurately by the digital video system made by us.
出处 《中华消化杂志》 CAS CSCD 北大核心 2005年第7期417-420,共4页 Chinese Journal of Digestion
  • 相关文献

参考文献5

  • 1Gertsch PH, Meister JJ. Pressure measurement in oesophageal varices: preliminary report on a new non-invasive method. Gut,1987,28: 1162-1165.
  • 2许建明,张磊.食管静脉曲张内镜气囊测压的临床价值[J].中华消化内镜杂志,2002,19(2):74-76. 被引量:8
  • 3张磊,许建明,孔德润,王玲,谢惠君,胡乃中.食管曲张静脉内镜气囊测压法的应用研究[J].中华消化内镜杂志,2004,21(1):36-39. 被引量:8
  • 4Gertsch PH, Wheatley AM, Maibach R, et al. Experimental evaluation of an endoscopic balloon for manometry of esophageal varices. Gastroenterology, 1991,101:1692-1700.
  • 5Pontes JM, Leitao MC, Portela F, et al. Endosonographic Doppler-guided manometry of esophageal varices: experimental validation and clinical feasibility. Endoscopy, 2002,34: 966-972.

二级参考文献3

  • 1梁扩寰 李绍白.门脉高压症(第1版)[M].北京:人民军医出版社,1999.128-135.
  • 2黄自平 梁扩寰 见: 梁扩寰 主编.门静脉高压症[A].见: 梁扩寰,主编.肝脏病学.第1版[C].北京:人民卫生出版社,1995.675-685.
  • 3许建明,张磊.食管静脉曲张内镜气囊测压的临床价值[J].中华消化内镜杂志,2002,19(2):74-76. 被引量:8

共引文献13

同被引文献38

  • 1吴建锋,黄挚雄,罗安,朱炼.静补型无创食管曲张静脉测压仪的设计[J].计算机测量与控制,2005,13(7):737-739. 被引量:1
  • 2谢岳,傅忠谦,孔德润,何兵兵.基于图像处理的无创性食管内静脉曲张压测量系统[J].生物医学工程研究,2007,26(1):24-28. 被引量:1
  • 3孔德润,许建明,付忠谦,何兵兵,孙斌,张磊,谢岳.基于计算机视觉的无创性食管曲张静脉测压仪[J].生物医学工程学杂志,2007,24(3):519-521. 被引量:2
  • 4Gertsch PH, Meister JJ. Pressure measurement in oesophageal varices:prelimiuary report on a new non-invasive method [J]. Gut, 1987,28(9) :1162 -5.
  • 5Xu J M, Kong D R,Fu Z Q, et al. Experimental Study of Esophageal Varix Manometrieal Digital Video System[ J]. J Gastroenterol Hepatol,2004,19 (Suppl) A638.
  • 6Kong DR, Xu JM, Zhang L, et al. A computerized endoscopic balloon manometry to detect esophageal varieeal pressure [ J ]. Endoscopy, 2009,41 (6) :415 -20.
  • 7Gertsch PH, Wheatley AM, Maibach R, et al. Experimental evaluation of an endoscopic balloon for manometry of esophageal varices [ J ]. Gastroenterology , 1991,101 (6) : 1692 - 700.
  • 8Gertsch PH, Meister JJ. Pressure measurement in oesophageal varices: preliminary report on a new non-invasive method. Gut, 1987, 28(9) : 1162 -1165.
  • 9Miller ES, Kim JK, Gandehok J, et al. A new device for measuring esophageal variceal pressure. Gastrointest Endosc, 2002, 56:284-291.
  • 10Pontes JM, Leit? o MC, Portela F, et al. Endosonographic Dopplerguided manometry of esophageal varices: experimental validation and clinical feasibility. Endoscopy, 2002, 34 : 966 - 972.

引证文献3

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部