期刊文献+

ADVIA120血液分析仪过氧化物酶鞘液试剂的研制与应用 被引量:5

Preparation and application of self-made peroxidase sheath reagent for ADVIA 120 hemocyte analyzer
下载PDF
导出
摘要 目的探讨自配过氧化物酶鞘液试剂是否能应用于ADVIA120血液分析仪。方法分析原装试剂成分、理化指标;研制效果相同的试剂,并作一系列比对试验。结果自配试剂的各项检测指标:N(%)、L(%)、M(%)、E(%)、B(%)、LUC(%)、WBCP等与进口试剂的检测结果基本一致,无显著性差异。结论自配的过氧化物酶鞘液试剂可替代进口试剂用于ADVIA120血液分析仪。 Objective To investigate if the self-made peroxidase sheath reagent can be used to substitute for PEROX SHEATH in ADVIA120 hemocyte analyzer. Methods The components and the physical and chemical parameters of original reagent of ADVIA 120 hemocyte analyzer were analyzed, and then the similar reagent was prepared. A series of comparisons between the self-made and original reagents were observed. Results All the parameters, e. g. , Neut ( % ), Lym ( % ), Mono ( % ), Eos ( % ), Baso ( % ), LUC ( % ) and WBCP,were analyzed by using the self-made reagent, and the results were similar to those analyzed by the original reagent. No significant difference between the two reagents was found and satisfactory results were achieved during three-month clinical trial. Conclusions The self-made peroxidase sheath reagent can be used to replace the imported original reagent PEROX for ADVIA 120 hemocyte analyzer.
出处 《临床检验杂志》 CAS CSCD 北大核心 2005年第5期350-352,共3页 Chinese Journal of Clinical Laboratory Science
基金 国家高技术研究发展计划(863计划)课题(2002AA2Z341B)。
关键词 血液分析仪 试剂 鞘液 hemocyte analyzer reagent research sheath liquid
  • 相关文献

参考文献3

二级参考文献8

  • 1段丹丽 熊英 赵兴梅.自配血细胞分析仪试剂除微粒方法探讨.四川省卫生管理干部学院学报,1997,16(2):121-121.
  • 2李国芳.CELL-DYN 900型血细胞分析仪使用国产试剂代替进口试剂的观察.中华血液学杂志,1988,19(9):561-561.
  • 3[1]International Committee for Standardization in Hematology;Expert p anel on cytometry.Reference method for the enumeration of erythrocytes and leuk ocytes[J]. Clin Lab Heamat,1994,16∶131-138.
  • 4[2]International Committee for Standardization in Hematology;Expert panel o n hemoglobinometry.Recommendations for reference method for hemoglobinometry in human blood(ICSH standard 1986)and specifications for international hemoglobincy anide reference preparation 3rd[J].Clin Lab Haematol,1987,9∶73-79.
  • 5[3]NCCLS.Precedure for Determining Packed Cell Volume by the Microhematioc r it Method-Second Edition;Approved Standard[S].NCCLS Document H7-A2, Vol.1 3 No 9,1993,13(9).
  • 6周洪华,颜箫,高航云,揭克敏,蔡尔泉,叶雯,万腊根,陈人骏.溶血剂对白细胞形态的影响[J].中华医学检验杂志,1997,20(2):87-89. 被引量:24
  • 7林发全,莫武宁.医学检验用稀释液的研制概述[J].陕西医学检验,2000,15(2):64-64. 被引量:10
  • 8张金生,黄天禄,李胜发.AC-920自动血细胞分析仪溶血剂的研制和应用[J].重庆医科大学学报,2000,25(1):73-76. 被引量:3

共引文献40

同被引文献23

  • 1李山,秦雪,黄军强,林发全.自配试剂在COULTER流式细胞仪检测HLA-B27中的应用探讨[J].广西预防医学,2005,11(3):164-165. 被引量:3
  • 2任津敏,赵淑会.用油镜进行白细胞分类的临床价值[J].职业与健康,2005,21(11):1836-1836. 被引量:2
  • 3徐宁,罗强,曹永坚,黄宪章,庄俊华,张秀明,梁伟雄.日立7170A生化检测系统检测项目的溯源研究[J].国际检验医学杂志,2006,27(12):1065-1067. 被引量:7
  • 4ADVIA120操作说明书.
  • 5丛玉隆,王丁.当代检验分析与临床[M].北京:中国科学技术出版社,2002:9-10,44-45.
  • 6陶义训,吴文俊,吴翊群.现代检验仪器导论[M].上海:上海科学技术出版社,2003:3-12.
  • 7NCCLS. Evaluation of precision performance of qualitative measurement methods. Approved Guideline-Second Edition [M ]. EP5-A2, 2004,24(25) : 1-38.
  • 8NCCLS. Method comparison and bias estimation using patient samples Approved Guideline-Second Edition. EP9-A2 [ M]. 2002, 22(19) : 1-52.
  • 9NCCLS. Evaluation of the linearity of quantitative measurement procedures: a statistical approach. Approved Guideline. EP6-A [M]. 2003,23( 16): 1-46.
  • 10Hitachi high-technologies corporation. Electrolyte analyzer accessory for model 7170, Instruction Manual.1-5.

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部