期刊文献+

CO_2监测分析仪在监测过程中对桥室温度的恒温控制

The Isothermal Regulation of Carbon Dioxide Intelligent Monitor
下载PDF
导出
摘要 CO2监测分析仪在监测过程中,CO2监测值的精度是由CO2监测分析仪的桥室参比温度是否恒定而决定的。老式监测分析仪桥室温度是采用断续的双位恒温调节,其温度调节误差大,一般只能在±1℃左右,所以CO2监测值的精度低。采用微机交流调压的CO2监测分析仪桥室恒温调节,其温度调节器采用PID调节,使桥室温度误差可控制在±0.2℃,提高了CO2监测值的监测精度。 The monitoring accuracy of intelligent carbon dioxide monitoring device depends on the effect of isothercommand of the carbon d er of the conducting cell. During monitoring if the temperature of the conducting cell could not maintain the fluctuation of temperature results in the unstability of the isothermal parameter. It directly influcuracy of monitoring carbon dioxide. The adoption of the duty ratio of alternation voltage under the computer would satisfy the demand of the isothermal control and increases the monitoring accuracy of ioxide monitor.
作者 李季渊
出处 《太原科技大学学报》 2005年第3期213-215,共3页 Journal of Taiyuan University of Science and Technology
关键词 微机调压 桥室恒温 CO2精度 voltage regulation by computer, isothermal conductiong cell,accuracy of carbon dioxide monitor
  • 相关文献

参考文献3

  • 1Coffron J W,Long W E. Practical interfacing techniques for microprocessor systems[M].New Jersey:Prenlice-Hall,2002.190-210.
  • 2王秀玲.A/D、D/A转换接口技术及数据采集系统设计[M].北京:清华大学出版社,2002.122-129.
  • 3张克彦.单片机这用程序设计[M].北京:北京航空航天大学出版社,2004.105-130.