期刊文献+

改进毛细管粘度计的新思路

An Improvement Method of the Capillary Viscosimeter
下载PDF
导出
摘要 本文详细分析毛细管粘度计样液中剪变率的分布规律,揭示毛细管粘度计测出血液的粘度是等效粘度,提出以管两端压强差△p为毛细管粘度计测出的血液表观粘度的比较参照条件,从理论上解决了ηa-△p曲线的测绘,提供了改进毛细管粘度计的思路.总之,对毛细管粘度计进行了具体深入、系统的探讨. This article analyses the shear rate distribution law among sample liquid of capillary viscosimeter in detail, proclaims the viscosity of blood tested by the capillary viseosimeter is equivalent, brings up that the pressvre disparity △p of two ends of tube as blood apparent viseosity's comparative reference eondition tested by capillary viseosimeter and settles ηa-△pcurre's surrey in terms of theory, provides the thinking of improving eapillary viseosimeter. In a word, goes on a particular,deeply and systematic inquire to the capillary viseosimeter.
出处 《中国医学物理学杂志》 CSCD 2005年第5期663-665,共3页 Chinese Journal of Medical Physics
关键词 毛细管粘度计 剪变率分布 等效粘度 粘度曲线 capillary viscosimeter shear rate distribution equiralent viscosity viscosity curve