摘要
目的:研究SPA-ELISA弓形虫病诊断试剂盒不同稀释液及灭活处理对阴性血清的保存效果。方法:正常人阴性血清(n=5)分别用0.01MPBS缓冲液、1%胎牛血清、15%甘油1∶100进行稀释后保存于4℃,分别于稀释后1d、1周、2周、4周、2个月、3个月用SPA-ELISA检测阴性血清OD410,比较其变化情况;同时观察灭活对阴性血清的影响。结果:用15%甘油稀释的阴性血清可以保存至3个月,用0.01MPBS缓冲液和1%胎牛血清稀释的阴性对照一个月之后保存效果较差。结论:15%甘油稀释的阴性血清可以比较长期稳定的保存,保证了SPA-ELISA弓形虫病诊断试剂盒的稳定性。
Objective: To investigate the storing effects of negative serum of diagnostic kit of Toxoplasma gondii, after diluted with different solutions and inactivation. Methods: Negative human serum samples (n=5) diluted 1:100 with 0.01 M PBS buffer, 1% BSA and 15% glycerol respectively were tested by SPA-ELISA at 1 day, 1 week,2 weeks, 4 weeks, 2 months, 3 months. The changes and difference of OD410 of them were compared. The effects of inactivation of serum were compared with that with no inactivation. Results: It showed that negative serums diluted with 15% glycerol can be well stored until 3 months. Comparatively, Negative serums kept unstable in 0.01M PBS buffer and 1% BSA after 1 month. Conclusion: Negative serum can be stably stored in 15% glycerol in long term, which ensures the stability of SPA-ELISA diagnostic kit of Toxoplasmosis.
出处
《天津医科大学学报》
2005年第3期365-366,共2页
Journal of Tianjin Medical University