期刊文献+

附加物对材料低温比热测量精度的影响

An Analysis of the Influence of Appendage on the Measurement Accuracy of the Specific Heat of Materials at Low Temperatare
下载PDF
导出
摘要 介绍了真空绝热法测量材料低温比热的原理及方法。考虑到测量系统中样品架、加热器及温度计等附加物引入的热容因素,建立了比热测量误差的传递关系式,并给出了分析数据。对具体测量中有关精度的问题进行了探讨。 The principle and method for measuring the specific heat of materials at low temperature with the vacumm adibatic method are described. The relationship of the measurement error transfer with the heat capacity of the appendage consisting of the specimen assembly rack, heater and thermometer in the apparatus is established and analysed data are given.Some problems about the measurement accuracy of the specific heat in the measurement system are discussed.
作者 李昭
机构地区 航天工业总公司
出处 《低温工程》 CAS CSCD 1996年第1期38-42,共5页 Cryogenics
关键词 比热 测量误差 温度计 材料 附加物 specific heat measurement error thermometer
  • 相关文献

参考文献4

  • 1张开达,钱寅虎,缪柏财,陈冠龙,霍焱.绝热法测量YBa_2Cu_3O_(7-δ)的比热[J].低温与超导,1989(3):59-62. 被引量:2
  • 2朱贤,冀勇夫,党震,韩军,白品贤,韩源洲,孙崇赤.固体低温比热的测定与分析[J]低温物理,1980(03).
  • 3谭志诚,周立幸,尹安学,陈淑霞,孙毅,王先良.80~370K精密自动绝热量热计及α-氧化铝和正庚烷的热容[J]化学学报,1980(06).
  • 4罗南星.测量误差及数据处理[M]计量出版社,1984.