PNP溶液检测自动生化分析仪精密度的方法研究

Precision measurement of antomatic biochemical analyzer with PNP solution
下载PDF
导出
摘要 目的:建立对硝基酚(PNP)溶液检测自动生化分析仪精密度方法。方法:通过酸性PNP溶液吸收峰扫描,在日立 7170自动生化分析仪上建立程序进行精密度观察,并对测定结果进行评价。结果:在酸性环境下PNP溶液最大吸收峰为 295-340nm;加入PNP溶液后的反应曲线是典型的终点比色法反应方式;连续记录30个工作日批内不精密度均未超过 1.20%,平均不精密度为0.55%;日间不精密度为2.22%。结论:PNP溶液检测自动生化分析仪精密度,方法简便,成本低 廉,能有效地反映自动生化分析仪工作状态。 Objective To set up an assay determining the precision of automatic analyzer with p-nitropbenol (PNP) solution. Methods Absorption peak scanning of PNP dissolved in acidic distilled water at pH 5.9 was performed, and the measuring program was established to determine the precision of Hitachi 7170 antomatic biochemical analyzer, and then the results were evaluated. Results The maximum absorption peak of PNP solution at pH5.9 was from 295nm to 340nm. The reactive curve accorded with end-point colorimetry, Imprecision was represented by CV value. During 30 working days, each CV value was less than 1.20%, and the mean value was 0.55% and the value from day to day was 2.22%. Conclusion The method for measuring the precision of automatic biochemical analyzer with PNP solution is simple, low-cost and effective.
出处 《医疗卫生装备》 CAS 2005年第11期15-16,共2页 Chinese Medical Equipment Journal
关键词 对硝基酚(PNP) pH指示剂 自动生化分析仪 精密度 PNP pH indicator automatic analyzer precision
  • 相关文献

参考文献6

二级参考文献26

  • 1[1]McCleary BV, McNally M, Monaghan D, et al. Measurement of alpha-amylase activity in white wheat flour, milled malt, and microbial enzyme preparations, using the Ceralpha assay: Collaborative study[J]. AOAC Int, 2002; 85(5):1096-1102.
  • 2[3]Kayamori Y, Katayama Y. Enzymatic method for assaying calcium in serum and urine with porcine pancreatic α-Amylase[J]. Clin Chem, 1994;40(5): 781-784.
  • 3[5]Hagishita T, Nishikawa M, Hatanaka T. A spectrophotometric assay for the transphos- phatidylation activity of phospholipase D enzyme[J]. Anal Biochem, 1999; 276(2):161-165.
  • 4Fujita T, Kawakami Y, Kohda S, et al. Assay of magnesium in serum and urine with use of only one enzyme, isocitrate dehydrogenase(NADP + ) [J]. Clin Chem, 1995,41:1302
  • 5Wenk RE, Lustagarten JA, Pappas NJ, et al. Serum chloride analysis, bromide detection, and the diagnosis of bromism[J].Am J Clin Pathol, 1976,65: 49
  • 6Randell EW, St Louis P. Interference in glucose and other clinical chemistry assays by thiocyanate and cyanide in a patient treated with nitroprusside [J]. Clin Chem, 1996,42: 449
  • 7Wang T, Diamandis EP, Lane A, et al. Variable selectivity of the Hitachi chemistry analyzer chloride ion - selective electrode toward interfering ions [ J ]. Clin Biochem, 1994,27: 37
  • 8Nagamine Y, Hamai Y, Chikamori K, et al. Asymptomatic hyperbromidaemia detected as pseudohyperchloridaemia measured with an ion selective electrode meter[J] .Scand J Clin Lab Invest, 1988,48:177
  • 9Mori L, Waldhuber S.Salicylate interference with the Roche Cobas Integra chloride electrode[J].Clin Chem, 1997,43:1249
  • 10Abdel-Wareth LO, Lirenman DS, Halstead AC, et al. Spurious rise in total carbon dioxide and chloride with negative anion gap after cystogram[J]. Pediatr Nephrol, 1995,9:348

共引文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部