根管长度电测仪测定后牙根管工作长度的临床研究 被引量:16

A clinical study of posterior teeth root canal working length measured by electronic apex locator
原文传递
导出
摘要 目的:探讨RootZX根管长度电测仪测量后牙根管工作长度的临床效果。方法:采用RootZX对60例79个后牙186个根管进行根管长度测量,以此长度进行根管预备,按测量长度插入牙胶尖摄X线牙片,计算出实际工作长度,计算RootZX测量根管工作长度的准确率。结果:在实际工作长度±0.5mm范围内,ZootZX测量准确率为87.6%,测量活髓和死髓牙根管工作长度的准确率无显著性差异(P>0.05)。结论:ZootZX根管长度测定仪测定根管工作长度准确性较高。 Objective: To investigate the clinical effect of Root ZX electronic apex locator in measuring posterior teeth root canal working length. Method: 186 root canals working length of 60 patients with 79 posterior teeth were measured by Root ZX and calculate by X-ray. Result: An accurate rate of real working length ranged ± 0.5 mm was 87.6%. The difference between vital pulp and necrotic pulp was not significant( P 〉0.05). Concluion: Root ZX is accurate in root canal working length measuring.
作者 吴国英
出处 《临床口腔医学杂志》 2005年第12期742-743,共2页 Journal of Clinical Stomatology
关键词 根管长度测量仪 后牙 工作长度 electronic apex locator posterior teeth working length
  • 相关文献

参考文献5

  • 1Martinez-Lozano MA, Fomer-Navarro L, Sanehez-Cortes JL, et al.Methodological considerations in the determination of working length[J]. Int Endod J, 2001,34(5) :371 - 376.
  • 2Dunlap CA, Remeikis NA, BeGole EA, et al. An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals[J]. J Endod, 1998, 24(1):48-50.
  • 3Shabahang S, Coon WWY, Gluskin AH. An in vivo evaluation of Root ZX electronic apex loeator[J]. J Endod, 1996,22:616- 618.
  • 4Ricard O, Roux D, Bourdeau L, et al. Clinical evaluation of the aceuraeyof the Evident RCM Mark lI apex locator [J]. J Endod.1991,17(11) : 567 - 569.
  • 5Shanhrokh S, William WY, Gluskin AH. An in vivo evaluation of mot ZX electronic apex locator[J]. J Endod. 1996,22(11):616-618.

同被引文献90

引证文献16

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部