期刊文献+

IP性能测试方法及使用寿命的研究 被引量:2

Study a test method of performance and using life of IP
下载PDF
导出
摘要 目的研究适合临床的IP性能测试方法并探讨CRIP实际使用寿命。方法根据CR系统影像处理“四象限”理论,调整相关象限参数,去除后处理影响,使整个CR系统如同屏/片组合。利用这种IP性能测试方法对20块IP进行铝梯及线对卡测试,所得数据进行统计学分析,以判断IP使用寿命。结果在万次曝光范围内,使用不同次数的IP性能无显著性差异(P>0.05);IP淘汰原因主要是机械划伤、异物擦伤。结论在CR运行过程中应注重使用环境、机械维护、使用方法的重要性,以延长IP使用寿命,减少成本消耗。 Objective To study a test method of performance for IP that is suitable for clinical practice as well as to evaluate the actual using life of CR IP.Method According to the "four quadrant theory" in image processing of CR system, we adjusted relevant quadrant parameters, removed the influence of post-processing, to make the CR system as a assemble screen/ film. Using this method, we used aluminum ladder and bar phantoms to test 20 pieces of [P, and analyzed the data statistically to evaluate the using life of [P.Result Under the limit of 10, 000 exposures, the function of different [P with different used times, was no significant different (P〉0.05).The main reasons for [P failure were mechanical injury and foreign body scratch.Conclusion It is worth to emphasize the importance of using environment, mechanical protection and using method in CR running, which will do good to prolong the using life of IP and decrease cost consumption.
作者 翟伟 陈治安
出处 《医疗设备信息》 2005年第12期9-10,共2页 Information of Medical Equipment
关键词 性能测试 成像板 计算机X线摄影 使用寿命 CR IP performance test image plate computer radiography
  • 相关文献

同被引文献17

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部