摘要 介绍了生物传感器技术测定血乳酸含量的原理和方法学评价.测定了50例健康儿童血乳酸含量的参考值为该仪器测定血乳酸的批内CV为1.27~2.10%,回收率为99.2~102.3%,线性范围为0~30mmol/L,只有快速,准确,自动化高,特异性强等优点。 his study introdticed the fundamentals of the immobilized enzyme biosensor technique in detection of plasmalactate and evaluated this techinique methodologically. 50 normal children's plasma lactate was tested.Results showedthat the normal plasmd lactate reference level was Coefficient of variation 1.27-2.10%,rate of recovery 99.2-102.3%,linearity range 0-3.0 mmol/ L were obtained within the same hatch with this in-strument.It has manifested such advantages as quiekness,accuracy,high specificity as well as process automation.
出处 《重庆医科大学学报》 CAS CSCD 1996年第1期43-45,共3页 Journal of Chongqing Medical University
关键词 固化酶 生物传感器 乳酸测定 血液 Immobilized enzyme Biosensensor Plasma lactate