生物组织热导率的脉冲-衰减法测量 被引量:1

Measurement of the tissue thermal conductivity by thermal pulse-decay method
下载PDF
导出
摘要 采用脉冲-衰减法测量生物组织的热导率,通过合理设计测量电路、选择电路参数,可以使脉冲输入期间探头输入功率的变化小于1%.当测量工况变化时,探头输入功率会产生较大变化,为此,提出了改进的脉冲衰减法.采用本方法,还可以适当提高脉冲输入期间探头的输入功率以增加探头温升,从而减少探头输入功率测量误差以及温度测量段温度测量误差的影响,提高热导率的测量精度.测量结果表明,由于脉冲-衰减法在推导时没有考虑探头尺寸的影响,造成热导率的测量值偏大;为此,进一步给出了一种热导率的测量值的简单修正方法,经修正后的热导率的最大误差为4.25%,平均误差为2.05%. The thermal pulse-decay method was used to measure the tissue thermal conductivity. The change of the probe bead supplied power can be less then 1% by properly designing the measurement of circuit and the selection of circuit parameters. When the measurement condition was changed, the probe bead supplied power would be changed greatly. An improved thermal pulse-decay method was presented to solve this problem. By using it, the probe bead supplied power can also be increased to raise the probe bead temperature. Thus the errors in the measured probe bead supplied power and the bead temperature can be reduced, and the measurement precision of the thermal conductivity can be enhanced. The measurement results showed that the measured tissue thermal conductivity was consistently larger than the corresponding standard value, and a simple corrected method was given to di- minish the measured error in the tissue thermal conductivity. After the correction, the maximum error of the measured thermal conductivity was 4.25%, and its average error was 2.05%.
作者 杨昆 刘伟
出处 《华中科技大学学报(自然科学版)》 EI CAS CSCD 北大核心 2006年第1期11-13,35,共4页 Journal of Huazhong University of Science and Technology(Natural Science Edition)
基金 中国博士后科学基金资助项目(2004036436)
关键词 热导率 生物组织 脉冲-衰减法 热疗 thermal conductivity bio-tissue thermal pulse-decay method hyperthermia
  • 相关文献

参考文献9

  • 1Chen M M, Holmes K R, Rupinskas. Pulse-decay method for measuring the thermal conductivity of living tissue[J]. Journal of Biomechanical Engineering,1981, 103: 253-260.
  • 2Arkin H, Holmes K R, Chen M M. A sensitivity analysis of the thermal pulse decay method for measurement of local tissue conductivity and blood perfusion [J]. Journal of Biomechanical Engineering,1986, 108: 54-58.
  • 3Arkin H, Xu LX, Holmes KR. Recent developments in modeling heat transfer in blood perfused tissues[J]. IEEE Transactions on Biomedical Engineering,1994, 41: 97-107.
  • 4Martin G T, Haddad M G, Cravalho E G, et al.Thermal model for the local microwave hyperthermia treatment of benign prostatic hyperplasia[J]. IEEE Transactions on Biomedical Engineering, 1992, 39: 836-844.
  • 5Charny C K, Levin R L. A three-dimensional thermal and electromagnetic model of whole limb heating with a MAPA[J]. IEEE Transactions on Biomedical Engineering, 1991, 38:1 030-1 039.
  • 6Tharp H S, Roemer R B. Optimal power deposition with finite sized, Planar Hyperthmia applicator arrays[J]. IEEE Transactions on Biomedical Engineering,1992, 39:569 579.
  • 7Tompkins D T, Vanderby R, Klein S A, et al.Effects of interseed spacing, tissue perfusion, thermoseed temperatures and catheters in ferromagnetic hyperthermia., results from simulations using finite element models of thermoseeds and catheters[J]. IEEE Transactions on Biomedical Engineering, 1994, 41:975-985.
  • 8Diederieh C J, Burdette E C. Transurethral ultrasound array for prostate thermal therapy: initial studies[J]. IEEE Transactions on Ultrasonics Ferroelectrics, and Frequency Control, 1996, 43; 1 011-1022.
  • 9Patel P A, Valvano J W, Pearce J A, et al. A self-heated thermistor technique to measure effective thermal properties from the tissue surface[J]. Journal of Biomechanical Engineering, 19879 109:330 335.

同被引文献3

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部