γ射线探测器能量响应标定技术 被引量:6

Technique of Calibrating Energy Response of γ-ray Detector
下载PDF
导出
摘要 叙述了γ射线探测器能量响应的标定原理和方法,即利用Compton散射将强^60Co源的1.25MeV γ射线转换为0.36~1.02MeV(25°~90°)范围内任意能量的系列单能γ射线;采用辐射屏蔽技术和提高能量分辨率方法对7射线探测器进行了该能区的能量响应标定;通过MCNP程序对探测器的能量沉积趋势和不同散射角散射7射线的能量分布进行模拟计算。结果表明:该标定系统信噪比达到了20:1左右,能量分辨率约为6%,0.66MeV的灵敏度与标准^137Cs源直照标定的灵敏度基本一致。 The paper introduces a method and principle for calibrating energy response of γ-ray detector. The energy response may be obtained by means of transforming γ-ray from ^60Co into a series of mono-energetic γ-ray in the range of 0.36~1.02 MeV, shielding radiation-interference and improving energy resolution. The energy distributions of scattered γ-ray at different angles and corresponding energy deposited in detector were simulated by MCNP code. The results show that the signal-to-noise of the system is about 20 : 1, the energy resolution is about 6%, and the sensitivity of detector to 0.66 MeV γ-ray is in agreement with the one to the γ-ray from ^137Cs.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2006年第1期88-91,共4页 Atomic Energy Science and Technology
基金 中国工程物理研究院预先研究资助项目(4210303)
关键词 能量响应 单能 能量分辨率 信噪比 energy response mono energy energy resolutions slgnal-to-noise