摘要
对该仪器进行评价使用RANDOX定值质控血清、高浓度病人血清对培Architectci8200生化免疫联体机作精密度、准备度、线性和样品携带率的评价测定。各项目在仪器不同分析系统中检测的日内CV为(0.4~0.8)%,日间CV为(0.5~0.85)%,与RANDOX定值质控血清的相对误差为(0.26~1.20)%。仪器线性良好,样品携带污染率低。适合大批量生化免疫样本的检测。
To comprehend the fuction of Abbott Laboratories Architect ci8200 biochemical immunity conjuctedly machine and evaluate the instrument. Evaluate the precision, accuracy, linear and sample carrying rate of the instrument by adopting RANDOX definite value serum and high density patient serum. The CV value of the same items measured in diferent analytic system is (0.4-0.8)%, the CV of values between deferent days is(0.5-0.85)%.The relative error to RANDOX definite value serum is (0.26-1.20)%.The instument has a good linearity and low sample carrying rate. Abbott Laboratories Architect Ci8200 biochemical immunity conjuctedly machine are well designed and has good performance, which is suitable for the measuring the sample in enormous quantities.
出处
《医疗设备信息》
2006年第1期65-67,共3页
Information of Medical Equipment
关键词
全自动
生化免疫联体机
性能
评价
automatic biochemical Immunity conjuctedly machine
performance
evaluation