摘要
用火焰原子吸收光谱法测定泛酸钙中的钙含量。以氯化钾溶液(0.1%,m/V)配制泛酸钙检测液,可消除空气-乙炔火焰的电离干扰作用。方法的线性范围为0.50~4.00μg/mL(r=0.9992),加标平均回收率为98.7%,RSD为0.46%(n=9)。
Abstract:Calcium in calcium pantothenate were determined by flame atomic absorption spectrophotometry. Ionization interferences carl be eliminated by the addition of 0. 1% potassium chloride(m/V) to samples and standard resolution. The calibration curve for calcium was linear over the range of 0.50 - 4.00 μg/mL. The average recovery was 98.7%, and RSD was 0.46% ( n =9).
出处
《中国兽药杂志》
2006年第2期17-18,共2页
Chinese Journal of Veterinary Drug
关键词
火焰原子吸收光谱法
泛酸钙
钙
flame atomic absorption spectmphotometry
calcium pantothenate
calcium