摘要
[目的]评价金属框架重建枕颈部的稳定性,为其临床应用提供力学依据。[方法]在7具完整成人枕骨颈椎(C0-6)标本上对横韧带切除齿状突切骨后失稳的(C0-6)节段,分别安装金属框架和棘突枕颈钢板内固定,通过摄像将1.53Nm载荷下的三维运动记录脊柱三维运动稳定性评价方法,评价其重建枕颈部即时稳定性的效果,[结果]在1.53Nm载荷作用下,金属框架组(C0-6)节段的前屈、后伸、侧弯和轴向旋转的ROM分别为5.9°、7.7°、5.6°、11.2°,比棘突枕颈钢板组分别缩小了157.6%、68.8%、58.9%、131.3%:[结论]枕颈框架组较有效地恢复枕颈部失稳后的即时稳定性,是枕颈融合术中较可靠的内固定方法。
[ Objective ] To provide the mechanical evidence of self manufactured metallic-rectangular frame on the reconstruction of the stahilit.v of occipitocervical junction. [ Method] Seven of the adult cervical spinal specimens which were entire wetcadavers( from occipitale to C6 ) were made to simulate the C0-2 fracture dislocation. These of specimens dislocated were fixed by metallic-rectangular frame and Occipito-cervical plate respectively. The three dimensional movements of C0-2 were recorded through photogrametry with a pure moment of 1.53 Nm. The range of motion(ROM) of each specimens in two fixation was caculated. [ Result] Loaded by 1.53Nm. in metallic-rectangular frame fixation, the ROM in flexion, extension, lateral bending and axial rotation were 5.9°,7.7°,5.6° , 11.2° respectively, decreased by 157.6% ,68.8% ,58.9% , 131.3% , compared with oceipitocervical plate/ixation group. [ Conclusion] Metallic-retangular frame fixation can reconstruct the stability of Oceipitocervicat junction immediaiev, and is a relativity effective internal fixation for occipitocervical fusion.