期刊文献+

影响实用化实时测温系统测温精度的若干因素 被引量:1

Several affecting factors of temperature measurement accuracy for a practical temperature measurement system
下载PDF
导出
摘要 在介绍一种利用钽酸锂热释电探测器实现的实用化双波长光纤测温仪的基础上,着重讨论了反射辐射、探测器本身的辐射与环境温度、光路中选择性吸收气体的光谱吸收及除此之外的其他诸种因素对系统测温精度的影响,并提出了相应的抑制措施. The effect of reflection radiation, radiation of the detector itself, the surrounding temperature, me selective absorption gas in optical path, and other factors on temperature measurement accuracy for a practical temperature measurement system were discussed based on a practical dual-wavelength fiber-optic colorimeter using LiTaO3 pyroelectric detector.
出处 《信阳师范学院学报(自然科学版)》 CAS 北大核心 2005年第4期397-399,419,共4页 Journal of Xinyang Normal University(Natural Science Edition)
基金 河南省自然科学基金项目(031400300204140030030511014300)
关键词 实时测温系统 测温精度 影响因素 practical temperature measurement system temperature measurement accuracy affecting factor
  • 相关文献

参考文献9

  • 1KHAN M A,ALLEMAND C,EAGAR T W.Noncontact temperature measurement Ⅰ:interpolation based techniques[J]. Rev Sci Instrum,1991, 62(2):392-402.
  • 2HAHN J W, RHEE C.Reference wavelength method for a two-color pyrometer[J].Appl Opt, 1987, 26(24):5276-5279.
  • 3HUNTER G B,ALLEHAND C D,EAGER T W.Multiwavelength pyrometer[J]. Opt Eng, 1985, 24(6): 1081-1085.
  • 4DEWITT D P.Inferring temperature from optical radiation measurement[J].Opt Eng, 1986, 25(4):596-601.
  • 5KRIVTZUM V,GRASS B,HERGENRODER R,et al.Temperature measurement of liquids by differential absorption of two diode laser: application of contactless optical detection in isotachophoresis[J].Appl Spectro, 2001, 55(9):1251-1258.
  • 6TAMURA Y, TATSUWAKI M, SUGIMURA T, et al.Temperature measurement of steel in the furnace[A].In:"Temperature, Its Measurement and Control in Science and Industry"[C].New York:American Institute of Physics, 1982.505-512.
  • 7施德恒,刘玉芳,陈玉科,朱遵略,孙金锋.一种高精度光纤测温系统工作波长的优化设计[J].光学学报,2004,24(4):542-547. 被引量:6
  • 8施德恒,刘玉芳,孙金锋,黄国庆,陈玉科.反射辐射和探测器本身的辐射对实时测温系统测温精度的影响及其抑制[J].中国激光,2004,31(1):105-110. 被引量:10
  • 9COATES P B.Analytic estimation of systematic errors in photoelectric pyrometry[J].High Temperatures-High Pressures.1985,17(6):507-518.

二级参考文献15

  • 1M. A. Khan, C. Allemand, T. W. Eagar. Noncontact temperature measurement. Ⅰ. Interpolation based techniques[J]. Rev. Sci. Instrum., 1991, 62(2):392-402.
  • 2J. W. Hahn, C. Rhee. Reference wavelength method for atwo color pyrometer [J]. Appl. Opt., 1987, 26(24):5276-5279.
  • 3G. B. Hunter, C. D. Allemand, T. W. Eagar.Multiwavelength pyrometer: an improved method [J]. Opt.Eng., 1985, 24(6):1081-1085.
  • 4D. P. DeWitt. Inferring temperature from optical radiation measurements [J]. Opt. Eng. , 1986, 25(4):596-601.
  • 5V. Krivtzum, B. Grab, R. Hergenroder et al.. Temperature measurement of liquids by differential absorption of two diodelaser: Application of contactless optical detection in isotachophoresis [J]. Appl. Spectroscopy, 2001, 55(9):1251.
  • 6C. H. Fan, J. P. Longtin. Laser-based measurement of temperature or concentration change at liquid surfaces [J].Journal of Heat Transfer, Transactions ASME, 2000, 122(4) :757-762.
  • 7H. H. Kee, G. P. Lees, T. P. Newson. Technique for measuring distributed temperature with 35-cm spatial resolution utilizing the Landau-Placzek ratio [J]. IEEE Photon. Technol.Lett. , 2000, 12(7) :873-875.
  • 8胡瑞华,栾松,孙宁克.发射率和温度同时测定的辐射测温仪[J].红外技术,1998,20(2):21-22. 被引量:17
  • 9辛军,冯其波,林铁生,王勇,姚敏言,廖延彪.高精度双波长非接触光纤高温计的波长最优设计[J].中国激光,1999,26(2):121-126. 被引量:22
  • 10王玮,周邦全,张在宣,吴孝彪,郭宁,余向东,王其良.分布型光纤拉曼光子温度传感器系统的测温精度[J].光学学报,1999,19(1):100-105. 被引量:27

共引文献13

同被引文献7

引证文献1

二级引证文献3

相关作者