Nafion修饰玻碳电极伏安法测定痕量铟 被引量:9

Voltammetric determination of In(Ⅲ) using a Nafion modified glassy carbon electrode
下载PDF
导出
摘要 报道了一种用Nafion修饰玻碳电极测定痕量铟的新方法.研究了Na-fion膜的有关特性和测定的条件,当富集时间为5 min时,峰电流与In(Ⅲ)浓度在1×10^-9~1×10^-7mol/L的范围呈良好的线性关系,检出限为1.46×10^-10mol/L.该法用于实际水样中痕量铟的测定,平均回收率为98.4%. This paper reports an application which a Nation modified glassy carbon electrode is used to determine trace Indium by Voltammetry. The experimental conditions have been optimized. The peak current is proportional to the concentration of Indium in the range of 1 × 10^-9 - 1 × 10^-7 mol/L, and the limit of detection is 1.46 × 10^-10 mol/L. The method has been applied to the determination of Indium in water samples and the average recovery is 98.4 %.
出处 《分析试验室》 CAS CSCD 北大核心 2006年第2期19-21,共3页 Chinese Journal of Analysis Laboratory
关键词 NAFION修饰电极 伏安法 Nation-modified electrode Voltammetry Indium
  • 相关文献

参考文献4

二级参考文献4

  • 1Mohan Katyal,Shri Prakash.Talanta[J],1977,24:367.
  • 2Laviron E.J.Electroanal.Chem.[J],1974,52:355.
  • 3安镜如,陈曦,谢玫,徐强,陈杨.铟-8-羟基喹啉络合物的极谱研究[J]分析试验室,1988(02).
  • 4李南强,章建康.铟-茜素配位剂极谱配位吸附波的研究[J]化学学报,1984(11).