期刊文献+

双刀口扫描法测量可见光CCD调制传递函数 被引量:3

Modulation Transfer Function Measurement of Charge-coupled Devices Using Double Scanning Knife
下载PDF
导出
摘要 介绍了一种新型的刀口扫描法测量可见光CCD的调制传递函数(MTF),首次采用1∶1的无像差光学系统,消除了测量光学系统MTF所带来的较大不确定度分量;采用双刀口方法和动态定焦扫描法,提高了测量效率和精度。 This paper introduces a novel method of the modulation transfer function (MTF) measurement of charge-coupled devices using the scanning knife. Firstly using the 1;1 free-aberration optical imaging system, the uncertainty effect of the optical imaging system MTF measurement is eliminated. The double knife is used to improve the efficiency and accuracy of measurement. A theory of dynamic focusing is established to improve the accuracy of the MTF measurement.
作者 王恒飞
出处 《计测技术》 2006年第1期44-47,共4页 Metrology & Measurement Technology
关键词 调制传递函数 CCD 双刀口 MTF charge coupled device (CCD) double knife
  • 相关文献

参考文献5

  • 1王庆有.CCD应用技术[M].天津:天津大学出版社,1991.
  • 2Mike Marchywka,Dennis G Socker.Modulation transfer function measurement technique for small-pixel detectors[J].Optical Engineering,1993,32(2):95-100.
  • 3Miles N Wernick,G Michael Morris.Effect of spatial coherence on knife-edge measurements of detector modulation transfer function[J].Oplical Engineering,1991,30(9):1394-1398.
  • 4GB/T4315.2-1998,光学传递函数测量导则[S].
  • 5John C Feltz,Mohammad A Karim.Modulation transfer function of charge-coupled devices[J].Application Optical,1990,27(2):717-722.

同被引文献16

引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部