期刊文献+

传导性角膜成形术后眼内压的变化(英文)

Measurements of intraocular pressure after conductive keratoplasty
下载PDF
导出
摘要 目的:探讨传导性角膜成形术(CK)对眼压测量可能影响。方法:在术前和术后第1wk,第4wk,分别用非接触式眼压计(NCT)、Goldmann眼压计(GAT)和Schiotz眼压计测量接受CK治疗的20例36眼的眼压,所得结果采用SPSS11.5软件进行统计分析。结果:和术前比较,CK术后NCT、GAT以及Schiotz眼压计测量值显著降低。眼压变化和年龄、性别、角膜曲率以及拟矫正屈光度无关。结论:CK术后压平式眼压计测量眼压值偏低。 AIM: To investigate the possible impact of conductive keratoplasty(CK)on intraocular pressure( IOP ). METHODS: 36 eyes of 20 patients with hyperopia underwent IOP measurement with non-contact tonometer (NCT), Goldmann applanation tonometer (GAT) and Schiotz tonometer before and after CK. RESULTS: After CK, a significant decrease was observed by NCT, GAT and Schiotz measurements. The change in IOP readings was not correlated with age, sex, keratometric readings, or attempted correction. CONCLUSION: The applanation tonometer atend to underestimate the true IOP after CK.
出处 《国际眼科杂志》 CAS 2006年第1期22-24,共3页 International Eye Science
关键词 眼压 传导性角膜成形术 intraocular pressure conductivekeratoplasty