期刊文献+

阳极溶出伏安法同时测定痕量锡和镉 被引量:10

Simultaneous Determination of Trace Tin and Cadmium by Anodic Stripping Voltammetry
下载PDF
导出
摘要 在HCl-NH_4Cl底液中,电解时间60s、电镀时间5s、除氧30s的条件下,Sn和Cd有稳定的溶出峰。Sn和Cd的检出限分别为6.5μg/L和3.0μg/L,测定范围分别为10~200μg/L和5~180μg/L。对含Sn和Cd为20μg/L的试液测定10份,RSD分别为5.61%和4.83%。方法用于实际样品测定,结果与文献方法相符。 In a supporting electrolyte HCl-NH_4Cl,the peaks of tin and cadmium were excellently separated and high sensitivities were achieved. The detection limits of Sn and Cd were 6. 5 μg/L and 3. 0 μg/L respectively,and the determination ranges were 10~200 μg/L and 5~180 μg/L respectively. The RSDs(n=10) as tested on 20 μg/L Sn and Cd samples were 5. 61% and 4. 83% respectively. The method has been applied to the analysis of real samples, and the results obtained were in agreement with those obtained from the existing methods.
出处 《岩矿测试》 CAS CSCD 北大核心 1996年第1期77-79,共3页 Rock and Mineral Analysis
关键词 阳极溶出法 同时测定 anodic stripping voltammetry,simultaneous determination, tin,cadmium