期刊文献+

仲钨酸铵萃余液中三氧化钨的流动注射测定 被引量:1

FIA-PHOTOMETRIC DETERMINATION OF WO3 IN THE RAFFINATE OF AMMONIUM PARATUNGSTATE PROCESSING
下载PDF
导出
摘要 基于在强盐酸介质及磷酸存在下钨被三氯化钛还原成钨蓝的性质,建立了流动注射测定仲钨酸铵生产过程苹余液中WO_3的简便新方法。测定波长为630nm,线性范围为0.01~0.1g·L^(-1),采样频率为60样·h^(-1)。对含0.044g·L^(-1)WO_3的典型试样连续13次测定,所得相对标准偏差为2.2%。 Based on the formation of a blue complex of tungstate after reduced by TiCl3 in a medium of hydrochloric acid and phosphoric acid,a FIA-photometric method for the determination of WO3 in the raffinate of ammonium paratungstate (APT) processing is proposed. Beer' s law is obeyed in the concentration range of 0.01-0.1g ?L-1 WO3. Samples can be analyzed with a frequency of 60 samples/h. For a typical sample containing 0. 044g ?L-1 of WO3,a relative standard deviation (RSD) of 2.2% is obtained for 13 measurements.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 1996年第3期143-144,共2页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 萃余液 流动注射分析 光度法 三氧化钨 Tungsten Raffinate Flow Injection Analysis (FIA) Spectrophotometry