导出
摘要 本试验采用50℃干烤法在不同实验条件下对A群脑膜炎球菌多糖菌苗的固体总量进行了测定,并对其结果进行了比较。实验结果表明:(1)1m1,2ml和4m1取样量的固体总量之间无显著性差异。(2)采用2ml取样量用不同规格称量瓶和不同规格吸管对测定结果无明显影响。(3)采用1m1、2ml和4ml取样量试验结果重复性均好。但从标准差及变异系数看,取样2ml或4ml者,试验误差更小,结果更为准确。 The dry weight of meningococcal polysaccharide group A vaccine were detected by 50℃ drying method,and the results were compared under the different conditions.The results showed: (1) there were not significance differentia between the test results of 1ml 、2ml and 4ml of sampling;(2)there were not influence on the dry weight of sample with different type of weighing bottle and pipette and (3) there were better repeatability on the results of 1ml、 2ml and 4ml of sampling.According to standard deviation and coefficient of variation of the test,we propose that the volume of sampling is 2ml or 4ml.
作者 朱莉萍
出处 《微生物学免疫学进展》 1996年第2期21-23,共3页 Progress In Microbiology and Immunology
关键词 A群脑膜炎球菌 多糖菌苗 固体总量 生物制品 Meningococcal polysaccharide group A vaccine Dry weight