期刊文献+

草酸氧化酶电极的研制及流动法测定草酸 被引量:3

Enzyme Electrode Based on Oxalate Oxidase with a Flow-cell for Oxalate Determination
下载PDF
导出
摘要 研制成灵敏度很高的具有流通池的草酸氧化酶电极,比较了三种结构不同的氧电极,选择了其中具有φ6.5mm阴极的一种。由达到稳态的电流测定,氧电流的降低与草酸浓度在2×10^(-6)mol/L~4×10^(-4)mol∥L范围内成线性关系。 A highly sensitive enzyme electrode with a flow-cell for oxalate was constructed. It is based on the immobilization of the enzyme oxalate oxidase via glutaraldehyde crosslinking on the tip of oxygen electrode. Three kinds of oxygen electrode with different construction were compared and one of them with a (?)6.5mm cathode was selected. The current change of oxygen reduction was measured. A calibration curve -obtained at steady state for the assay of oxalic acid showed linearity of current changes vs. oxalate concentration between 2 ×10-5-4×10-4mol/L.
出处 《北京大学学报(自然科学版)》 CAS CSCD 北大核心 1990年第2期141-146,共6页 Acta Scientiarum Naturalium Universitatis Pekinensis
关键词 草酸 测定 氧化酶 电极 氧电极 尿 Oxalate oxidase Oxygen electrode Membrane Flow-cell Steady current
  • 相关文献

参考文献4

  • 1华惠珍,化学传感器,1987年,7卷,4期,11页
  • 2张荣坤,化学传感器,1987年,7卷,4期,11页
  • 3董绍俊,医学生物电化学方法,1983年
  • 4冯强生,化学学报,1981年,39卷,1期,55页