摘要
使用自动活检装置在B超引导下经皮肾活检60例,结果表明:在B超定位、导向保证穿刺准确、安全的前提下,用自动活检枪弹时较细的18G切割针所取标本长度和含肾小球个数与传统的15GMenghini型粗针负压抽吸活检组20例对照无显著性差异(P>0.05,P=0.75),一次取材成功率高达93.3%(P<0.001);98.3%可满足病理诊断;仅发生轻微并发症8.3%(5/60)。穿刺后观察分级指标与对照组进行秩和检验(u=2.14,P=0.03),前者优于后者。
cases of percutaneous automatic renal biopsy guided by sonography wereanalyzed in this paperThe results showed that the first puncture successful rate in the group us-ing the automatic biopsy gun with a rather small 18 gauge cutting needle had greater advantageover that of the aspiration group using a 15 gauge large-core(P<0001)There was no signifi- cant differance in the mean length of the tissue core(P>0.05)and the average number ofglomeruli(P= 0.75)obtained by these two methods.The specimens obtained from automaticbiopsy group were adequate for histologic diagnosis purpose(98.3%),and there was only a rath-er low incidence of complications.
出处
《广西医科大学学报》
CAS
1996年第1期62-64,共3页
Journal of Guangxi Medical University
关键词
肾活检
自动活检术
超声波诊断
renal biopsy
guided by B-ultrasound
automatic biopsy gun