摘要
目的:评估国产微型钛钉种植支抗即刻负载稳定性。方法:随机选择犬下颌骨的一侧为实验组即加力组,另一侧为对照组即不加力组,植入国产微型钛钉各两颗。实验组钛钉间施力200克。三个月后处死动物,分别测量实验组、对照组钛钉间的位移变化及种植体—骨界面剪切力。结果:实验组钛钉间位移平均减少0.47mm;对照组几无变化;实验组种植体-骨界面剪切力明显大于对照组,平均为367.42牛顿。结论:国产微型钛钉作为种植支抗可满足临床要求。
Objective: To evaluate the stability of homemade titanic miniscrew as an Orthodontic anchorage loaded immediately. Methods: 16 homemade titanic miniscrews were inserted in the baccal posterior region of the lower jaw of four dogs. The control group (no force loaded) and the experimental group (200g- forces loaded immediately) were divided randomly. The changes of distances between the miniscrews of both groups were measured respectively after 3 - month force application period. The shear bone strength of the interface between miniscrew and bone in each group was also measured with a push test.Results: The mean changes of distances between 2 miniscrews in the experimental group was 0.47mm and nearly 0.00mm in the control group; the shear bone strength of the experimental group, with an average level of 367. 42N, was significantly higher than the control group one. Conclusion: The homemade titanic miniscrew can serve as an orthodontic anchorage loaded immediately.
出处
《中国口腔种植学杂志》
2006年第1期13-16,共4页
Chinese Journal of Oral Implantology
基金
深圳市科技局资金资助项目(0310)
关键词
微型钛钉
支抗
剪切强度
Miniscrew
Anchorage
Shear bone strength