摘要 目的:探讨圆柱形螺纹种植体螺纹形态对骨组织应力分布的影响,为临床设计和选择最佳的螺纹形态提供理论依据。方法:建立了12种包含不同螺纹形态种植体的颌骨骨块三维有限元模型:1 mm、2 mm和3 mm矩形螺纹设计(S-1、S-2和S-3);90°、60°和30°V形螺纹设计(V-1、V-2和V-3);45°、30°和15°支撑形螺纹设计(B-1、B-2和B-3);45°、30°和15°反支撑形螺纹设计(R-1、R-2和R-3)。对所有模型进行应力分布和Von M ises应力峰值的比较。结果:垂直向加载下S-2、V-3、B-3、R-1、R-2和R-3螺纹表现出较好的应力分布状态,在颊舌向加载下S-1、S-2、V-3、B-3、R-2和R-3螺纹表现出较好的应力分布状态。结论:提示S-2、V-3、B-3、R-2和R-3螺纹设计均适用于圆柱形种植体,反支撑螺纹为圆柱形种植体的最佳螺纹设计选择。 Objective: To determine the optimal thread shape for an experimental cylinder implant. Methods: Twelve three - dimensional finite element models of different thread shapes were created : 1 mm, 2mm, and 3mm square designs ( S - 1, S - 2, and S - 3);90°, 60°, and 30° V - shaped designs ( V - 1, V - 2, and V - 3) ; 45°, 30°, and 15° buttress designs ( B - 1, B - 2, and B - 3) ; 45°, 30°, and 15° reverse buttress designs ( R - 1, R - 2, andR - 3). The stress distributions of all models were compared under axial and buccolingual loads. Results: Models S - 2, V - 3, B - 3, R - 1, R - 2, and R - 3 showed better stress distribution than other models under axial load, and models S - 1, S - 2, V - 3, B -3, R -2, and R -3 showed better stress distribution than other models under buccolingual load. Conclusion: S -2, V - 3, B - 3, R - 2, and R - 3 thread shapes all appear to be suitable for use in a cylinder implant, and reverse buttress is the optimal design in a cylinder implant.
出处 《口腔医学研究》 CAS CSCD 2006年第2期155-158,共4页 Journal of Oral Science Research
关键词 种植体 三维有限元分析 力学加载 应力分布 螺纹设计 Dental implant Three -dimensional finite element analysis Load Stress distribution Thread design
  • 相关文献

参考文献17

  • 1Murat C,Joke D,Michel DC,et al.Implant design and interface force transfer a photoelastic and strain-gauge analysis[J].Clin Oral Impl Res,2004,15:249-257
  • 2Ivanoff CJ,Grondahl K,Sennerby L,et al.Influence of variations in implant diameters:a 3-to 5-year retrospective clinical report[J].Int J Oral Maxillofac Implants,1999,14:173-180
  • 3Brunski JB.Biomechanical considerations in dental implant design[J].Int J Oral Implantol,1988,5:31-34
  • 4Bumgardner JD,Boring JG,Cooper RC Jr,et al.Preliminary evaluation of a new dental implant design in canine models[J].Implant Dent,2000,9:252-260
  • 5Misch CE.Contemporary Implant Dentistry,2nd ed.St.Louis:Mosby,1999
  • 6Branemark PI,Hansson BO,Adell R,et al.Osseointegrated implants in the treatment of the edentulous jaw:experience from a ten year period[J].Scand J Plast Reconstr Surg Suppl,1977,16:1-132
  • 7Misch CE,Bidez MW,Sharawy M.A bioengineered implant for a predetermined bone cellular response to loading forces.A literature review and case report[J].J Periodontol,2001,72:1 276-1 286
  • 8Buchs AU,Hahn J,Vassos DM.Interim clinical study report:a threaded,hydroxylapatite-coated implant-five year postrestoration safety and efficacy[J].J Oral Implantol,1995,21:266-274
  • 9Timoshenko SP,Goodier JN.Theory of Elasticity.McGraw-Hill International Book Co[J],Singapore,1984
  • 10Zienkiewicz OC.The Finite Element Method in Engineering Science,4th Edition[J].McGraw-Hill,New York,1989

同被引文献255

引证文献15

二级引证文献51

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部