摘要
建立了一种用电子捕获检测器气相色谱法(GC- ECD)测定大气及空气分离装置液空、液氧中N2O的测定方法.通过选择性条件试验,确定了用GC- ECD法测定大气及空气分离装置液空、液氧中N2O含量的最佳色谱柱为不锈钢填充柱,固定相为GDX-103(0.154~0.18 mm粒径);最佳载气流速为18 mL/min,最佳分离的柱温为30℃.方法具有较高的灵敏度,检出限为:当进样量为0.5mL时,可测准至N2O浓度0.1 μL/L的气体,检出限为0.014 μL/L.方法的准确度较高,以峰面积定量,其标准偏差为0.013;相对标准偏差为3.74%.
The method of gas chromatography-electron capture detection (GC-ECD) N2O in air, liquefied air and liquefied oxygen was studied and set up. The chromatography were chosen by a lot of tests . The best column is GDX-103 (0. 154 The rate of carrier gas is 18 mL/min. The temperature of column is 30℃. This method precision. The detection limit can reach up to 0. 014 μL,/L with a RSD of 3.7%. for the determination of optimum conditions of -0.18 mm pore size). has high sensitivity and
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
2006年第4期554-556,共3页
Chinese Journal of Analytical Chemistry
关键词
大气
液空
液氧
氧化亚氮
电子捕获检测器
热导池检测器