摘要
本文分析了心阻抗图圆台圆柱体组合模型理论存在的若干问题。指出:该理论在基本原理方面有明显错误;其计算心搏出量的公式并非科学推证的结果,而是人为拼凑的产物,公式的计算值不能反映被测个体的真实情况。按此理论及公式来作医学检测,有可能导致临床误诊。
The article analysed the theoretical defects and existing errors of the combination model of cir-cular truncat cone and cylinder about the impedance cardiogram,and indicated that the stroke volume formula of the model is not the result of scientific deduction,but the product of man-made gathering.Its calculation can notrellect the actual situation of the stroke volume measured.To use the formula on medical measurement will prob-ably result in clinical misjudge.
出处
《中国医学物理学杂志》
CSCD
1996年第2期65-67,共3页
Chinese Journal of Medical Physics
关键词
心阻抗图
组合模型
可靠性
Impedance cardiogram Combination model Reliability