摘要 A sigma-edged design of intraocular lens (IOL) is proposed to minimize the reflected glare images associated with the edge. The optimal sigma inverse edge is investigated when pupil diameter is 5.0 mm by the established Escudero-Sanz's wide angle model. The non-sequential ray tracing program of Zemax-EE (Zemax Development Corp., San Diego, USA) is used to investigate the sigma edge of reducing the potential for edge glare phenomena. The results show that sigma-edged design can significantly reduce the reflected glare intensity on retina if the angle of its inverse edge is taken from 20 to 70 degrees. A sigma-edged design of intraocular lens (IOL) is proposed to minimize the reflected glare images associated with the edge. The optimal sigma inverse edge is investigated when pupil diameter is 5.0 mm by the established Escudero-Sanz's wide angle model. The non-sequential ray tracing program of Zemax-EE (Zemax Development Corp., San Diego, USA) is used to investigate the sigma edge of reducing the potential for edge glare phenomena. The results show that sigma-edged design can significantly reduce the reflected glare intensity on retina if the angle of its inverse edge is taken from 20 to 70 degrees.
出处 《Chinese Optics Letters》 SCIE EI CAS CSCD 2006年第5期299-302,共4页 中国光学快报(英文版)
关键词 GLARE Health care Health risks Implants (surgical) REFLECTION Surgery Glare Health care Health risks Implants (surgical) Reflection Surgery