期刊文献+

吸附络合物体系的示波分析 被引量:4

Oscillographic Analysis for Adsorbable Complex System
下载PDF
导出
摘要 In(Ⅲ)在pH457HAcNaAc0.1%VC0.06%HCuP溶液中,用示波计时电位法可获得良好示波图,其切口电位为E=090(Vs.SCE),切口高度与In(Ⅲ)浓度在800×10-7~14×10-5mol/L范围内成正比,检出限可达50×10-7mol/L吸附络合物组成为In(Ⅲ):铜铁试剂(HCup)=1∶1。 In this paper, a system containing adsorbable complex(0.2mol/L HAc-0.4mol/L NaAc-0.1%VC-0.06%HCup-In) was studied by oscillographic chronopotentiometry. Because the complex can be adsorbed on the surface of electrode. The higher sensitivity of determination can be used in trace analysis. In the experiment, a good oscillogram is obtained. Its potential of incision is 0.9(vs,SCE). The height of incision is propoftional to concentration of indium(Ⅲ) over the range of 8.00×10 -7 ~1 4×10 -5 mol/L and the limit of detection is 5.0×10 -7 mol/L.
机构地区 南京大学化学系
出处 《分析科学学报》 CAS CSCD 1996年第2期136-138,共3页 Journal of Analytical Science
基金 国家自然科学基金 河南省科学基金
关键词 吸附络合物 示波分析 铜铁试剂 Oscillographic Analysis Adsorbable Complex Indium(Ⅲ)