胸部X射线CT扫描中的放射防护最优化 被引量:1

OPTIMIZATION OF RADIATION PROTECTION IN X-RAY CT SCAN OF CHEST
下载PDF
导出
摘要 X射线螺旋CT机不同扫描条件下,用有机玻璃均匀模体测量多层扫描的CT剂量指数值,用均匀水模体测量CT值的均匀性及其噪声。比较了不同扫描条件下,患者肺部肿块影及片状影的CT图像质量。结果表明:曝露量为402、5、7.5 mAs时的CT剂量指数值分别不超过曝露量为115 mAs时的40%、30%和15%。本文用CT扫描“优化因子”作为图像质量、病人所受剂量的综合评价指标,实现了CT扫描条件优化的定量分析。虽然,随X射线曝露量减少,均匀模体断层内CT值噪声增加、正常图像百分率下降,但从图像质量、病人剂量综合考虑,X射线胸部CT扫描的曝露量宜取25 mAs。 Under different scan conditions, CT dose indices of multiple scan were measured using homogeneous PMMA phantom. Homogeneity and noise of CT values were also measured using a water phantom. CT image quality for mass and patch in the lung of patients that were selected randomly was compared. The results indicated: that CT dose indices for 40, 25,7.5 mAs are not more than 40%, 30%, 15 % of that for 115 mAs respectively. Quantitative analysis of scan conditions was carried out by using CT scan optimization factor as a comprehensive index to evaluate image quality and patient dose. The noise of CT values in homogeneous phantom increased and percentage of normal image decreased with reduction of X-ray exposure. Considering image quality and patient dose comprehensively, the optimized exposure for chest scanning is 25 mAs.
出处 《辐射防护》 CAS CSCD 北大核心 2006年第3期129-136,共8页 Radiation Protection
关键词 X射线CT 胸部扫描 曝露量 CT剂量指数 CT值均匀性 CT值噪声 CT扫描优化因子 X-my CT, Chest Scanning, Exposure, CT Dose Index, Homogeneity of CT Value, Noise of CT Value, CT Scan Optimization Factor
  • 相关文献

参考文献8

  • 1Naidich D P, Marshall C H, Gribbin C, et al. Low-dose CT of Lungs: Preliminary Observations. Radiology,1990. 175(3) :729
  • 2Kaneko M, Eguchi K, Ohmatsu H. Peripheral Lung Cancer: Screening and Detection with Low-dose Spiral CT Versus Radiography. Radiology, 1996.201 (3) : 798
  • 3Nitta N, Takahashi M, Murata K, et al. Ultra Low-dose Helical CT of the Chest. ATR, 1998. 171(2) : 383
  • 4聂永康,蔡祖龙,赵绍宏.低剂量螺旋CT筛选检查早期肺癌的初步研究[J].中华放射学杂志,2002,36(3):230-234. 被引量:122
  • 5Rusinek H, Nadich D P, Mc Guinness G, et al.Pulmonary Nodule Detection: Low-dose Versus Conventional CT. Radiology, 1998. 209(1) : 243
  • 6Nadich D P, Yankelevitz D F, Mc Guinness G, et al.Noncalcified Pulmonary Nodules Missed on Low-dose Helical CT. Radiology, 1999.213(2) :303
  • 7Henschke C L, Mc Cauley D L, Yankelevitz D F, et al.Probability of Malignacy of Small Noncalcified Nodules Detected on Low-dose CT of the Chest. Radiology, 1999.212(2) : 561
  • 8中华人民共和国国家计量检定规定.医用诊断计算断层摄影装置(CT)X射线辐射源.JJG 961-2001

二级参考文献8

  • 1Naidich DP,Marshall CH,Gribbin C,et al.Low-dose CT of the lungs:preliminary observations[].Radiology.1990
  • 2Sone S,Takashima S,Li F,et al.Mass screening for lung cancer with mobile spiral computed tomography scanner[].The Lancet.1998
  • 3Nesbitt JC,Putnam JB Jr,Walsh GL,et al.Survival in early-stage lung cancer[].The Annals of Thoracic Surgery.1995
  • 4Itoh S,Ikeda M,Isomura T,et al.Screening helical CT for mass screening of lung cancer: application of low-dose and single-breath-hold scanning[].Radiologia Medica.1998
  • 5Smith RA,Glynn TJ.Epidemiology of lung cancer[].Radiologic Clinics of North America.2000
  • 6Henschke CI,McCauley DI,Yankelevitz DF,et al.Early Lung Cancer Action Project: overall design and findings from baseline screening[].The Lancet.1999
  • 7Shah R,Sabanathan S,Richardson J,et al.Results of surgical treatment of stage I and Ⅱ lung cancer[].Journal of Thoracic and Cardiovascular Surgery.1996
  • 8Lam S,Shibuya H.Early diagnosis of lung cancer[].Clinics in Chest Medicine.1999

共引文献121

同被引文献1

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部