期刊文献+

固相萃取-离子色谱法测定乳与乳制品中可致癌物硝酸盐的研究 被引量:8

Determination of Carcinogen Nitrate in Milk and Dairy Products by Spe-Ion Chromatography
下载PDF
导出
摘要 研究了采用固相萃取-离子色谱法测定乳与乳制品中可致癌物硝酸盐。将制备好的样品加入适量乙酸沉淀蛋白,离心,取上清液以4mL/min的流速通过DIONEXOnGuardⅡRP固相萃取柱去除脂肪等疏水性化合物后,直接进样测定。用30mM的KOH溶液等度淋洗,流速为1.0mL/min,采用AS11-HC型阴离子色谱柱进行分析。在此检测条件下,硝酸盐有很好的线性(r=0.9999),回收率为85.5% ̄108%。实验结果表明,该方法简便、快速,且检测结果准确。 The method for determination of carcinogen nitrate in milk and dairy products by the SPE(Solid phase extration) - ion chromatography was researched on the basis of experiment.After added acetic acid,settled protein and centrifuged, the sample was passed though the DIONEX OnGuard Ⅱ RP sample pretreatment cartridge at a flow rate of 4mL/min to wipe off fat and other hydrophobic compound,then was determined directly.The separation was performed with 30mM KOH as isocatic mobile phase at a flow rate of 1.0mL/min and accomplished with a Dionex Ionpac AS11-HC column.The calibration curve of nitrate was linear.And the correlation coefficient was 0.9999,The recoveries were 85.5%-108%. The method was simple, rapid and accurate.
出处 《现代科学仪器》 2006年第2期71-73,共3页 Modern Scientific Instruments
关键词 固相萃取 离子色谱 乳与乳制品 硝酸盐 SPE ion chromatography milk and dairy products nitrate
  • 相关文献

参考文献5

  • 1GB/T5009.33-2003..食品卫生国家标准汇编(4)..,,....
  • 2Alonso,A.; Etxaniz,B.; Martinez,M.D.Food Additives and Contaminants.1992,9,111-117
  • 3Eggers,N.j.;Cattle,D.L.J.Chromatogr.1986,354,490-494
  • 4Jackson,P.E.;Haddad,P.R;Dilli,S.J.Chromatogr.1984.295,471-478
  • 5胡平,王淑惠,王军,赵伟.固相萃取-离子色谱法测定乳清粉中的硝酸盐[J].现代科学仪器,2005,22(6):82-84. 被引量:6

二级参考文献4

  • 1食品卫生国家标准汇编(4),GB/T5009.33-2003.
  • 2A.Alonso; B.Etxaniz; M.D.Martinez.Food Additives and Contaminants.1992,9,111-117.
  • 3N.J.Eggers; D.L.Cattle.J.Chromatogr.1986.354,490-494.
  • 4P.E.Jackson; P.R.Haddad; etc.Chromatogr.1984.295,471-478.

共引文献5

同被引文献741

引证文献8

二级引证文献149

相关作者