原子吸收光谱法测定水溶液中银不确定度的评定 被引量:10

EVALUATION OF UNCERTAINTY OF MEASUREMENT IN THE FAAS DETERMINATION OF SILVER IN AQUEOUS SOLUTION
下载PDF
导出
摘要 分析了影响原子吸收光谱法(AAS)测定水溶液中银的主要因素,并对其所引起的不确定度分量进行评估。对含银0.10~2.00 mg·L-1的水溶液,置信水平95%时,测量结果的扩展不确定度为0.020 mg·L-1,有效自由度为50。 The main factors, affecting the accuracy of the FAAS determination of silver in aqueous solution were discussed, and the uncertainties of measurements caused by these factors in the determination were evaluated. An expanded uncertainty of 0. 020 mg ·L^-1 , at 95% confidencel level and with an effective degree of freedom of 50, was obtained in the determination of 0.10 to 2.00 mg· L^-1 of silver.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2006年第5期385-387,共3页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 原子吸收光谱法 不确定度 Uncertainty of measurement FAAS, Silver