摘要
[目的]探讨简便快捷的耗氧量(OC)测定方法.[方法]将国标法(GB/T 5750-1985)的水浴加热30 min改为直火加热5 min,硫酸溶液浓度由(1+3)改为(1+4),测定水样耗氧量(OC),并与国标法进行比较分析.[结果]改进的测定方法可靠,RSD为1.75%~2.29%,配对t检验,P>0.05,与国标法比较差异无统计学意义.[结论]改进法操作简单,易于掌握,耗时短,缩短了检验周期,提高了工作效率,适合于基层实验室推广使用.
[Objective]To develop a simple and rapid method for determination of oxygen consumed(OC)in water. [Methods]The method was improved in the base of national standard method (GB/T 5750-1985), which involved in fire heating 5 min instead of water bath heating 30rain described in standard method, and using the concentration of sulfuric acid in water 1 +4 instead of 1 +3. The values of OC determined by two methods were compared and analyzed. [Results] The improved method was reliable,with relative standard deviation between 1.75% to 2.29%. The values determined by two methods hadno statistical difference with t-test ( P 〈0.05). [Conclusion]The improved method was simple, rapid,reliable,and could be popularized in grass-root labs.
出处
《预防医学论坛》
2006年第3期310-311,共2页
Preventive Medicine Tribune
关键词
耗氧量(OC)
国标法
改进法
加热条件
Oxygen consumed (OC)
National standard method
Improved method
Heating condition