期刊文献+

试剂携带污染对血清总胆汁酸测定结果的影响及消除措施 被引量:29

The influence of the reagent residual contamination for the result of determination of serum total bile acid and its elimination measure
下载PDF
导出
摘要 目的消除循环酶法测定血清总胆汁酸(TBA)时试剂携带污染对结果的干扰。方法循环酶法测定20份血清标本中的TBA含量,然后分别测定总胆固醇(TC)、三酰甘油(TG)、高密度脂蛋白胆固醇(HDL-C)等9个项目后再测定TBA含量。结果在测定TC、TG、HDL-C后测定TBA导致20份血清标本结果存在系统误差,分别为14.95、15.36、22.20μmol/L。3种干扰试剂组分中的TBA含量分别为1 016.9、1 061.2、1 216.8μmol/L。结论合理安排TBA与其他检测项目的分析顺序,或设置避免试剂携带污染功能程序,用碱性清洗液冲洗试剂针,可以有效地消除TC、TG、HDL-C试剂对血清TBA测定结果的干扰。 Objective To eliminate the interference of the reagent residual contamination for serum total bile acid (TBA) measurement. Methods The levels of TBA were determined in 20 serum samples by circulating enzymatic method. The TBA was determined after 9 items including total cholesterol( TC), triglyceride(TG) and high-density lipoprotein cholesterol(HDL-C). Results System error occurred in 20 serum samples when TBA were measured after TC, TG, HDL-C. They were 14.95, 15.36, 22.20 μmol/L. TBA concentration in three interference reagents were 1 016.9, 1 061.2, 1 216.8μmol/L, Conclusions To deal with TBA and other items analysis in good sequences, or to set programe of avoiding reagent residual contamination, or to wash reagent needles with alkali washing buffer that can eliminate the reagent residual contamination of the TC, TG and HDL-C.
出处 《检验医学》 CAS 北大核心 2006年第3期228-230,共3页 Laboratory Medicine
关键词 总胆汁酸 血清 酶循环法 携带污染 Total bile acid Serum Circulating enzymatic method Residual contamination
  • 相关文献

参考文献2

二级参考文献5

共引文献8

同被引文献79

引证文献29

二级引证文献129

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部