摘要 低压多丝正比室为放射性束核反应的靶前粒子提供位置和时间信号。本工作用241Amα源研究电压、工作气体压强等因素对低压多丝正比室位置分辨的影响。实验结果表明:在不同的气体压强条件下,低压多丝正比室的位置分辨均可达到1 mm,但考虑到时间、效率等因素,实验中需尽可能提高气压。 Low pressure multi-wire proportional chamber(LPMWPC) provides excellent position and time signals for the radioactive beam induced reactions. By using the ^241 Am α source, the influencing conditions on the position resolution, including the voltage and the working gas pressure(800-1 330 Pa), were studied. The best position resolutions at different gas pressures are less than 1 mm. Considered the time resolution and detection efficiency, it is necessary to make the gas pressure as high as possible.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2006年第3期338-341,共4页 Atomic Energy Science and Technology
基金 国家重点基础研究发展规划资助项目(G2000077400) 国家自然科学基金资助项目(10405002)
关键词 低压多丝正比室 放射性束 位置分辨 low pressure multi-wire proportional chambers radioactive beam position resolution
  • 相关文献

参考文献5

  • 1谭继廉,张金霞,吴伯冰,李惠林,靳根明.用于中能重离子的组合式充气探测器[J].核电子学与探测技术,1996,16(1):1-5. 被引量:5
  • 2OTTINI-HUSTACHE S,MAZUR S,AUGER C,et al.CATS,a low pressure multiwire proportionnal chamber for secondary beam tracking at GANIL[J].Nucl Instrum Methods,1999,A431:476-478.
  • 3ASSAMAGAN K,BARKER K,BAYATYAN G,et al.Time-zero fission-fragment detector based on low-pressure multiwire proportional chambers[J].Nucl Instrum Methods,1999,A426:405-419.
  • 4YE Y L,DI Z Y,LI Z H,et al.Study and application of low pressure multi-wire proportional chambers[J].Nucl Instrum Methods,2003,A515:718-824.
  • 5李智焕,陈陶,叶沿林,江栋兴,华辉,李湘庆,王全进,葛愉成,庞丹阳,狄振宇,谭继廉,张金霞,凤盈,张保国,肖志刚,王宏伟,吴和宇,靳根明.低气压多丝正比室及其在放射性束流实验中的应用[J].高能物理与核物理,2002,26(9):964-970. 被引量:4

二级参考文献1

共引文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部